P 26
Freely scalable differential pressure transducer; Measurement ranges between ± 10 Pa and ± 100 kPa.
In addition to differential pressure data, the P 26 differential pressure transducer also records positive and negative overpressures. Measurement ranges between 100 Pa and 100 kPa are freely scalable from 10 to 100%; includes option of switching to other units, such as mmH2O. Two optional contact points allow the user to connect visual or audible alarms that would be triggered any time values fall above or below limits. The instrument can, if necessary, communicate with a master control unit via an RS 232 interface. The instrument can be mounted either on a wall or on top-hat rails. This pressure transducer is especially suited for monitoring pressure in clean rooms, for use in special mechanical engineering applications, or for monitoring a sterile environment such as a laboratory, pharmaceutical plant or operating room.
The LCD is optional and can display results in pressure units other than Pa or kPa.
A freely adjustable time constant (up to 60s) allows the instrument to produce a calm output signal, even under fluctuating pressure conditions
Technical data: measurement ranges ±10 Pa to ±100 kPa freely scalable from 10 to 100% high level of accuracy and long-term stability automatic zero-point calibration eliminates zero-point drift low level of hysteresis high level of overpressure protection time constants and output characteristics can be configured (root-extracted / linear) can be mounted on top-hat rails Options: contact points with adjustable switching outputs RS 232 interface Accessories: configurator | |